ThePointe180impulse™isanISEElectrolyteAnalyzerspecificallydesignedforthedirectdeterminationofNa+,K+andCl-byionselectiveelectrodesinwholeblood,serum,andaqueousQCmaterial.
Reagentpacks,electrodes,andtheinstrumentitselfallcontainuniqueelectronicsallowingforselfcalibrationandmonitoringfortroubleshooting.
Theanalyzerisabottle-free,standalone,benchtop,ISEelectrolyteanalyzerdesignedforthedirectdeterminationofNa+,K+andCl-byionselectiveelectrodes.Theunitcontainsabuiltinthermaldotmatrixprinter.Theunitisprogrammablethroughcustomsoftwareallowingtheusertodefineprogramsfollowingtheirownestablishedprotocols.Thesoftwareisdisplayedonaninteractivetouchscreenwhichguidestheuserstep-by-stepthroughtheprocessusingonscreenprompts,errormessagesandgraphicaldisplays.
Reagentsaresuppliedinpackswhichplugintothesideoftheunitwithnousercontacttowastematerial.Theunitwilltrackreagentexpirationandquantityofreagentremaining.*
*NotavailableforclinicalhumanbloodtestingintheUS
Catalog#: | Product: |
90.P3900-10 | Pointe180impulse™ |
Catalog#: | Product: |
90.B11-100 | ReagentPack |
90.B11-110 | Na+Electrode |
90.B11-120 | K+Electrode |
90.B11-130 | Cl-Electrode |
90.B11-140 | ReferenceElectrode |
90.B11-310 | Tri-LevelQCKit |
Simplicity
IntelligentUserInterface
ConvenientandSafe
Method: | Directmeasurementbyionselectiveelectrode |
SampleType: | WholeBlood,Serum,AqueousQCMaterial |
SampleSize: | 60µl±5µl |
SampleRate: | 48/hourmaximumwithoutprinting |
| 45/hourmaximumwithprinting |
MeasurementRanges: | Sodium — 40-205mmol/L |
| Potassium — 1.5-15.0mmol/L |
| Chloride — 50-200mmol/L |
| |
SampleProbe: | 316StainlessSteel |
Vessels: | Capillarytubeswithadaptor,tubesupto100mmdepth,syringes,samplecups |
Electrodes: | Sodium(Na+)Electrode |
| Potassium(K+)Electrode |
| Chloride(Cl-)Electrode |
AnalysisTime: | Typically<60seconds |
PatientDataStorage: | 200results(withcapabilitytoassigna15characterpatientidentifier) |
QCDataStorage: | 270reADIngs(30readingseachof3controlsforeachofthe3electrodes,includesLevey-Jennings) |
Calibration: | Ondemandorautomaticfrom1to8hourintervals |
PowerRequirements: | 100-240VAC,50-60Hz90Wmaximum |
Dimensions: | 14”wx12”hx8”d(35.5cmx30.5cmx20cm) |
Weight(withreagentpack): | 14.5lbs(6.6kg) |
Display: | Interactivetouchscreen3.5”(9cm)LCDcolorgraphicdisplay |
Interface: | IndirectviaUSBDrive(Included) |
Printer: | Thermaldotmatrixwithgraphiccapability,29characterwidth |
ThePointe180isasmall,semi-automatedchemistryanalyzerwithsixfiltersandalargetestmenu.Ithasanonboardincubatorwith12wellsat37°C,atemperaturecontrolledcuvettewell,andinternaltemperaturecontrol.Theanalyzerleadstheuserthroughtestperformanceinastep-by-stepmanner.Thenewgraphicssoftwareallowsextendedtimedabsorbanceintervaldataandkineticdataplotstobeprinted.
Theanalyzerhas48programmedand21openchannels,andreads,calculates,andprintsresultsinlessthanthreeseconds.Theanalyzerreadstesttubesuponinsertion----noneedtopressabutton.OptimizedparametersforPointeScientificreagentsarealreadyentered-justselectthetestnumber.Openchannelsareavailableforprogrammingothertests.Thenormalrangeforeachtestisautomaticallyprintedout,andresultsthatareoutofthenormalrangeareflagged.Normalrangescanbeeditedtofittheareaorpopulation.ResultsgeneratedbythePointe180arecomparabletoresultsreportedbyhospitalandreferencelabs.
Theanalyzercanusesinglepointcalibrationbystandard/factor,multi-pointcalibrationwithpointtopointcurvefit,percentabsorbanceandratebystandardorfactor.Thesecalculationoptionsallowtheanalyzertoperformendpoint,initialrate,kineticrateandbatchkineticmethodologies.
Catalog#: |
Product: |
90.P5025-180 |
Pointe180 |
Catalog#: |
Product: |
90-12B-01 |
12x75mmtesttubes |
90-P1907-03 |
PrinterPaper |
BICHROMATICPHOTOMETERSYSTEM:
Differentialwavelengthstocorrectfortesttubeimperfectionsandtohelpremoveinterferingsubstances.
ON-BOARDINCUBATOR:
12wellsat37°C.
TEMPERATURECONTROLLEDCUVETTEWELL:
Constant37°C.Thisfeatureisnecessaryforperformingkineticassays.
INTERNALTEMPERATURECONTROL:
Theanalyzermonitorstheincubatorandcuvettewelltemperatures.Thetemperaturesareautomaticallycorrectedandtheactualtemperatureindegrees(°C)isdisplayedonthescreen.
LAMPSAVINGFEATURE:
Theinstrumentautomaticallyshutsoffthelampiftheinstrumenthasnotbeenusedfor15minutes.Thisfeaturecanpreservethelampforuptotwo(2)years,dependingonactualusage.
MAINTENANCEFREE:
Theinstrumentdoesnotrequireanyscheduledmaintenanceprocedures.
SELF-PROMPTINGOPERATION:
Theanalyzerleadstheuserthroughtestperformanceinastep-by-stepmanner.
TESTTUBEACTIVATEDREADING:
Theanalyzerreadstesttubesuponinsertion-noneedtopressabutton.
FASTREADOUT:
Theanalyzerreads,calculatesandprintsresultsinunderthree(3)seconds.
PREPROGRAMMEDCHEMISTRYPROCEDURES:
OptimizedparametersforPointeScientificreagentsarealreadyentered-justselectthetestnumber.Openchannelsareavailableforprogrammingothertests.
PREPROGRAMMEDDATAREDUCTION:
Calculationprogramsforcardiacriskfactors,glycohemoglobinandIron/TIBCarepartofthegeneraltestmenu.
PREPROGRAMMEDNORMALRANGES:
Theanalyzerprintsthenormalrangesandcanbeeditedtofittheareaorpopulation.
PREPROGRAMMEDFLAGS:
Theanalyzerwillflagresultsthatareoutofthenormalrangeorexceedthelinearlimitsoftheanalyzerorreagents.
AUTOMATICWAVELENGTHSELECTOR:
Theanalyzerautomaticallypositionsthepreselectedfiltersforeachtest:340,405,550,600nm.Theinclusionofthe340nmfilterenablestheanalyzertoperformassaysintheultra-violet(UV)mode,whichisthepreferredmodeforenzyme/kinetic,assays.
VARIETYOFCALCULATIONOPTIONS:
Theanalyzercanusesinglepointcalibrationbystandard/factor,multi-pointcalibrationwithpointtopointcurvefit,percentabsorbanceandratebystandardorfactor.Thesecalculationoptionsallowtheanalyzertoperformendpoint,initialrate,kineticrate,andbatchkineticmethodologies.
EIAPROCEDURES:
ThePointe180iscapableofperformingenzymeimmunoassayproceduressuchasTheophylline,TSH,TU,T4andothersbasedonitscurvefittingcalculationcapABIlity.
GLYCOHEMOGLOBINANALYZER:
Readsandcalculatesresultsautomaticallyin%glycohemoglobin.
CARDIACRISKANALYZER:
Automaticallycalculates%HDL,LDL/HDLratio,andassignsariskfactorusingdirectmeasurementsofHDL,LDL,andCholesterol.TheCardiacRiskAssessmentisbasedoneitherDirectLDLMeasurementorcalculatedLDL.
ELECTROLYTEANALYZER:
PerformsCO2andChlorideusingchemistrymethodologies.
IRON/TIBCANALYZER:
RunsTotalIronandUnsaturatedIronBindingCapacityandautomaticallycalculatesTotalIronBindingCapacity.
CHEMISTRYANALYZER:
Performsallbasicchemistryproceduresquickly,easilyandeconomically,usingthelatestmethodologies.ResultsgeneratedbythePointe180analyzerarecomparabletoresultsreportedoutbyhospitalandreferencelabs.
GRAPHICCAPABLESOFTWARE:
Newgraphicsoftwareallowsextendedtimedabsorbance,intervaldata,andkineticdataplotstobeprinted.
LinearMeasurementRange: |
0.0to2.50AbsorbanceUnits(A) |
PhotometricAccuracy: |
±(1%ofthereading+0.005A)at1.5A |
|
±(2%ofthereading+0.005A)at1.5-2A |
LightSource: |
TungstenHalogenLamp |
Wavelengths: |
340,405,450,490,550,600nm |
TypicalHalfBandpass: |
10nm |
TypeofFilter: |
Multiple-cavitysealedinterference |
TubeSize: |
12mmroundstandard |
MinimumFillVolume: |
1mL |
MaximumFillVolume: |
2mL(fortemperaturecontrolto37°Conly) |
Temperature: |
37°C(±0.3°),Blockstayson,Cellhason/offswitch |
Lampsavertime: |
After15minutesidle |
Speed: |
Reads,calculatesandprintsresultsin3seconds |
Display: |
Alphanumeric,16character,LCD |
Printer: |
Thermal,dotmatrix,20charactersperline |
Keyboard: |
16key,membraneswitch,4X4,audIBLeresponse |
Calculationmodes: |
Singlepointcalibrationbystandardorfactor |
|
multipointcalibrationwithpointtopointcurvefit |
|
multipointpercentabsorbanceratestandardorfactor(batchorsingle) |
|
fixedtimekinetic |
Fuserequirements: |
120VAC-0.6A,3AGsloblo |
Power: |
Switchselectablepowersupply(115Vor230Vindicated) |
|
VoltageSource:110-120/220-240VACfrom50to60Hz |
|
InstallationCategory:CATII |
|
Powerconsumption:45Watts |
|
Fuse:Two0.6A,Trating,250Vsloblo,3AGtypefuses |
|
Allpowercordsmustbeapprovedforthecountryofuse |
Dimensions: |
9.5X13.25X3.75inches(24X34X10cm) |
Weight: |
Approximately10lbs(4.5Kg) |